Instruction manual

C-5
Note
Function #
Function name Range
Default
Sq F-14
Batch Finish output
width
0.0 - 25.5 (sec.) 0.0
Auto
Sq F-15
Batch monitoring
timer
0 – 255 (sec.) 0
Auto
Sq F-16
Supplementary flow
open timer
0.01 - 2.55 (sec.) 0.10
Auto
Sq F-17
Supplementary flow
close timer
0.1 - 25.5 (sec.) 0.1
Auto
Sq F-18
Discharge start wait
timer
0.0 - 25.5 (sec.) 0.0
Auto
Sq F-19
Discharge valve
close wait timer
0.1 - 25.5 (sec.) 0.1
Auto
Sq F-20
Discharging time
monitor timer
0 - 255 (sec.) 0
Auto
Sq F-21 Add the final to zero band setting 0
0
Does not add the final
Loss
1
Adds the final
Sq F-22 Add the final to full setting 0
0
Does not add the final
Loss
1
Adds the final
Sq F-23 Batch start function key 0
0
Disable
1
[ESC]
2
[+/-]
Sq F-24 Emergency stop function key 0
0
Disable
1
[ESC]
2
[+/-]
3
[ON/OFF]
Sq F-25 Discharge start function key 0
0
Disable
1
[ESC]
2
[+/-]
Only one among Sq
F-23, -24, and –25
can be used. When
two or more are set,
the function
with a
higher priority
becomes available
and the others will be
disabled.
Priorities: Sq F-23 is
1st,
Sq F-24 2nd, and
Sq F-25 3rd.
Sq F-26 Automatic Tare at batch start 0
0
Disable
Nozzle
1
Not disabled
It tares automatically when detects the “zero ba
nd” signal at a
weighing completion. (ready to re-start)