Instruction manual

C-3
Function #
Function name
Default
FncF-10
Automatic return to the normal mode after indicating
set point data
1
0
Disable
1
Not disabled
When no operation from keys for more than 10 seconds, it returns to the
normal mode
.
FncF-11 Preset tare weight by code 1
0 Disable
Preset tare without code number can be recalled in spite of the
setting of
0.
1
Not disabled
( without tare-clear)
When there is no tare data to be recalled, previous t
are weight will be
used.
2
Not disabled
( with tare-clear)
When there is no tare data to be recalled, it clears the tare data.
FncF-12 Recalling set point data by code 1
0 Disable
1 Not disabled
FncF-13 Setting set point code by code 1
0 Disable
1 Not disabled
FncF-14 Set point code NOT to be recalled by code 0
0 None
1 Zero band
2 Full
3 Free fall
4 Zero band, Full
5 Zero band, Free fall
6 Full, Free fall
7 Zero band, Full, Free fall
FncF-15
Continuous command of accumulation / manual print
0
0 Not disabled
1 Disable on accumulation
2 Disable on manual print
3
Disable on both accumulation and
manual print
With this function disabled, command is accepted
only o
nce when the net weight exceeds zero band
value. After that, command will not be accepted
unless load on the scale is removed. This function
is to avoid double accumulation or printing for the
same data.