Instruction manual

A
-2
Interfaces
Input
(CONTROL IN)
6 points (6 bits, 2 Commons)
Dry contact or open collector
Signal is insulated from an internal circuit by an
optocouple
r.
Terminal functions selectable by the function settings
Control I/O
Output
(CONTROL
OUT)
6 points (2 bits, 1 Common x 3)
Relay contact output
AC current output available
Rating: 24 VDC, 0.5 A (Resist
ance load), 220 VAC, 0.5 A
(Resistance load)
Standard serial
output(C.LOOP OUT)
Output for connection to a peripheral device manufactured by A&D (0 to 20 mA
current loop signal)
Weighing Capabilities
Zero point correcting
capability (Zero)
Clears the gross to zero in compliance with a command from the ZERO
key
s, Control I/O, etc.
Capable of enabling or disabling oper
ations when unstable.
A battery backs up zero correction value.
Adjustable range: Can be freely set within 1 to 30 % of weighing capacity.
Zero tracking capability
Detects gross zero point drift and performs zero co
rrection automatically.
Tracking time: 0.0 to 5.0 (seconds)
The range can be set freely.
Tracking width: 0.0 to 4.5 (D). The range can be set freely
Tare subtracting
function
Clears the net weight to zero in compliance with a command from the
T
ARE key, Control I/O, etc.
Capable of enabling or disabling oper
ations when unstable or negative
weight.
The battery backs up the tare value.
Adjustable range: gross W
eighing capacity
Preset tare subtracting
function
Sets/recalls presets tare for each code number.
Adjustable range: 0 < gross W
eighing capacity
Stability detecting
capability
When a weighed value fluctuation amount per sampling enters a set wid
th
within a set time, the stability mark is illuminated, judging it a stable
condition.
Capable of confirming stability through Control Output.
Stability detection time: 0.1 to 5.0 (seconds).
The range can be set freely.
Stability detection width: 0 to 9 (D). The range can be set freely
Digital filter function
Integrating capability
Two digital filters connected in series
Cutoff frequency range: 0.7 to 11 (Hz)
Accumulation
capability
Accumulates the net weight automatically or manually.
The battery backs up an accumulated weight and the number of
accumulated weighing co
unts.
Accumulated weight range: 0 to 99999999
Accumulated weighing times range: 0 to 99999999