Instruction manual
A
-1
APPENDIX A: SPECIFICATIONS
Specifications
A/D Converter Block
Input sensitivity 0.3μV/D or more
Zero correction range 0 – 20mV (0 - 2mV/V)
Load cell excitation
10 VDC±5%, 120mA with remote sensing capability
Up to four 350Ω load
cells can be connected
Zero point
±(0.2μ+0.0008% of Dead Load)/ typical
Temperature
coefficient
Sensitivity ±0.0008%/°C typical
Non-linearity ±0.01% F.S.
Input noise ±0.3μVp-p or less
Maximum measurement
voltage
32mV(3.2mV/V)
A/D conversion method Sigma-delta method
A/D internal resolution Approximately 1,000,000 max.
Sampling speed 100times /second
Maximum display resolution 10000D (D: minimum division
Digital Block
Element
•
•
•
7 digits, 7-segment, Character height:13mm
Fluorescent display tube
Display color: Cobalt blue, Status display in orange
Main Display
Item
displ
ayed
• Gross weight, net weight
Element
•
•
•
Element: 8 digits, 7-segment, Character height: 7mm
Fluorescent display tube
Display color: Cobalt blue, Status display in orange
Sub Display
Item
displ
ayed
•
•
Tare, Final, Accumulated weight (Can be selected in the function
setting.)
Calibration and function setting information in a setting mode
Status Display section
• 14 “” marks
LED 5 LEDs (SET POINT, TOTAL, OVER, ACCEPT, UNDER)
Type A
[ZERO], [TARE], [NET/GROSS], [ON / OFF], [“7”/UNDER/ANF],
[“8”/OVER/CALF], [“9”/PRESET TARE], [SET POINT],
[“4”/OP.PRELIM/OUTF], [“5”/ZERO BAND/SIF], [“6”/FULL/RSF] ,
[FUNC.], [“1” / FINAL/FncF] , [“2” / FREE FALL/SQF] , [“3”/PRELIM/INF] ,
[ F ], [“0”/CODE] , [+ / -], [ESC], [ENTER]
Keys
Type B
[ZERO], [TARE], [NET/GROSS], [ON / OFF], [“7”/ANF], [“8”/CALF],
[“9” /PRESET TARE], [SET POINT], [“4” / Lo / OUTF],
[“5” / Lo-Lo / SIF],
[“6” /ZERO BAND/RSF] , [FUNC.], [“1”/Hi-Hi/FncF] , [“2”/Hi /SQF] ,
[“3”/Go/INF] , [ F ], [“0”/CODE] , [+ / -], [ESC], [ENTER]
Switches
•
•
•
Main power switch
Calibration disable switch (used when required by law.)
Located inside the case.