Instruction manual

8-14
Commands
* See {8-6 Data Transmitting Format —Data Format} for “Data Format”.
* See {8-7 Command Mode— Communication procedure} for “Procedure”.
Command
Command name
Function
Procedure
Data
Format
RW
(Request Weight)
Recalls the weight B A/B
The content of the response data is the same as that set with RSF-01.
Same operation as when the manual print key is pressed.
MZ (Make Zero) Zero clear A
MT (Make Tare) Subtracts the tare A
Subtracts the tare. Display shows the net weight.
MN (Make Net)
Changes from Gross to Net display
A
MG (Make Gross)
Changes from Net to Gross display
A
CT (Clear Tare) Tare clear A
AM (Acc.Memory) Accumulates weight/count A
CA (Cancel Acc.) Cancels accumulation A
Subtracts the previous weight/count from the accumulated data.
SSXX (Set Set points)
Sets set points to the code
designated by XX
C E
When no code designated (blank), set points now in use are set.
Set point data length: 7 bytes including polarity.
RSXX
(Request Set
point)
Recalls set points from the code
designated by XX
B E
When no code designated (blank), set points now in use are recalled.
CCXX (Change Code)
Recalls the code number
designated by XX
A
BB (Begin Batch) Begins batch A
This command is available only in the built-in automatic program mode weighing.
If used in other weighing mode, "IE" will be transmitted.
BD
(Begin
Discharged)
Begins discharge A
This command is available only in the built-in automatic program mode weighing.
If used in other weighing mode, "IE" will be transmitted.