Instruction manual
8-11
Format Type Format Name
Form E Set Point Setting Format
See {Appendix E: Data output examples}
Code:61
Final(Go):+1000
Free fall(N/A):-25
Preliminary
(Hi-Hi):+150
Optional preliminary
Lo-Lo
:+400
Over limitHi+12
Under limit(Lo+5
Zero band
+10
Full (N/A):+16000
Preset Tare
+1763
Terminator
• Type B model for the Check
weighing
• This format is available only in
command mode.
• Data length: 76 bytes ( 60 bytes)
• Set “SSXX” command ignoring decimal
point. Decimal point will cause an error.
• “RSXX” command ignores decimal
point.
61
+0001000
-0000025
+0000150
+0000400
+0000012
+0000005
+0000010
+0016000
+0001763
CR
LF
Form F Function Data Read-Out/Write-in Format
• This format is available only in the command mode.
• Function Data is read out from Basic functions to Calibration functions.
• “EOF (1Ah)” is output after recalling the last data.
Read-out Write-in
Function #
Setting Data
with polarity and
decimal point
Term.
Function #
Setting Data
with polarity and
decimal point
Term.
FNC01 , +12345678
C
R
L
F
E
O
F
FNC01
,
+12345678
C
R
L
F
Form G Code number Read-Out/Write-in Format
Code
Code #
61
Separator
Final
Final Setting value
Separator
Free Fall
Free Fall Setting
value
…
Setting value
Total number overflow
1s or 0s
Terminator
CODE
61
FINAL
+000100
F _FALL
-0000025
NO OF
000000000
CR
LF
EOF
• This format is available only in command
mode.
• Read-out and Write-in are performed in the
following order;
Final (FINAL), Free Fall (F FALL), Prelim
(PRELIM), Optional preliminary (OP
PRELIM), Over Limit (OVER), Under Limit
(UNDER), Full (FULL), Preset Tare (TARE),
Accumulated Weight (TTL WT),
Accumulated Number (TTL NO),
Accumulated Weight Overflow (TTL OF).
• “EOF (1Ah)” is output after recalling the last
data.
• Write-in is performed code by code.
• Overflow/ Underflow flag;
0: Not Overflow / Underflow
1: Overflow / Underflow