6-11 CALF-14=8 Weighing Mode
Check weighing 3)
Output signal Output condition
Zero band
Gross weight
Zero band
Hi-Hi Hi-Hi limit < Net weight
Hi Hi limit < Net weight
Go Lo limit
Net weight
Hi limit
Lo Net weight < Lo limit
Lo-Lo Net weight < Lo-Lo limit
When an output condition is established, a relevant output terminal is turned on (power continuity with the output
common) or off. “Full flow”, “Medium flow”, and ”Dribble flow”: OFF, the other outputs: ON.
Select an output terminal at the outF-01-06 setting. See {Appendix C: Function List}.
Signal is output only in the normal mode. No output in the calibration mode nor function setting mode.