Instruction manual

6-17
6-10 CALF-14=7 Weighing Mode
(Check weighing 2)
Output signal Output condition
Zero band
Gross weight
Zero band
Hi-Hi Target weight + Hi-Hi limit < Net weight
Hi Target weight + Hi limit < Net weight
Target weight + Hi-Hi limit
Go Target weight - Lo limit
Net weight
Target weight + Hi limit
Lo Target weight - Lo-Lo limit
Net weight < Target weight - Lo limit
Lo-Lo Net weight < Target weight - Lo-Lo limit
When an output condition is established, a relevant output terminal is turned on (power continuity with the output
common) or off. “Full flow”, “Medium flow”, and ”Dribble flow”: OFF, the other outputs: ON.
Select an output terminal at the outF-01-06 setting. See {Appendix C: Function List}
Signal is output only in the normal mode. No output in the calibration mode nor function setting mode.