4-3
4-4 Gravity compensation
1. Get in the calibration mode
Press [CALF] while pressing [ENTER] in the normal mode, and press [ENTER] next. “CAL
Set” is displayed. (ready to calibrate)
* Press [ESC] to return to the normal mode.
2. Get in the Gravity compensation mode
Press [FUNC.]. “GrAvity” is displayed on the main display and your “g” value is on the
sub-display..
* Press [ESC] to return to the normal mode.
3. Enter the know acceleration rate for your location
Refer to {Appendix D: Gravity acceleration map} and enter the value on the map.
Press [ENTER] to save the calibration data.
* Press [+/-] to cancel the input and re-input.
* Press [ESC] to return to the normal mode.