Instruction manual
4-2
Actual load calibration procedures
Calibration procedures vary by the type of calibration. Follow the step number in the table below to
calibrate.
* When
“C ErrXX” is displayed, there is something wrong with the calibration. See {Appendix
B: Error Codes}.
Zero calibration only
Span calibration only
Zero and Span calibration
Get in the calibration mode.
1 1 1
Press [CALF] while pressing [ENTER] in the normal mode, and press [ENTER] next.
“CAL Set” is displayed. (ready to calibrate)
* Press [ESC] to return to the normal mode.
Calibrate Zero.
2 2
Press [ENTER] to get into the zero calibration mode.
Main display:
“CAL 0” / Sub display: “ ”
* To browse the gross weight, press [SET POINT], it will be displayed on the sub
display. Press [SET POINT] to erase this display.
* Press [ESC] to return to the normal mode.
Press [ENTER] at a stable reading (“-” is on beneath “Stable”) with no load.
“-------” is displayed for 2 seconds and the displays show;
Main display:
“CAL SPn” / Sub display: “ Capacity weight ”
* To end calibration without zero calibrated, press [ESC],
“CAL End” is displayed.
Move on to Span calibration.
2
To skip zero calibration and perform span calibration only, press [ENTER]. When
“CAL 0” is displayed, press [F].
Calibrate Span.
3 3
The displays show;
Main display: “CAL SPn” / Sub display: “ Capacity weight ”
* The capacity weight is of the value set in CALF-04.
* To browse the gross weight, press [SET POINT], it will be displayed on the sub
display. Press [SET POINT] to erase this display.
* Press [ESC] to return to the normal mode.
* To end calibration without span calibrated, press [ESC],
“CAL End” is displayed.
Input the weight of the calibration mass using the numerical keys of ([0] - [9]).
Press [ENTER] at a stable reading (“-” is on beneath “Stable”).
“-------” is displayed for 2 seconds and the displays show;
Main display:
“CAL End” / Sub display: “ ”
End Calibration after saving the calibration data in EEPROM.
3 4 4
Press [ENTER] to save the calibration data in the EEPROM and return to the normal
mode.
* To return to the normal mode without saving the calibration data, press [ESC].