Instruction manual
C-15
Function #
Function name Range
Default
CALF-14 Weighing mode 3
1 Normal batching (Customer Programmed Control Mode)
2 Loss-in-weight (Customer Programmed control Mode)
3 Normal batching (Built-in automatic program mode)
4 Loss-in-weigh (Built-in automatic program mode)
5 Nozzle Controlled Weighing Mode (Built-in automatic program mode)
6 Check weighing 1
7 Check weighing 2
8 Check weighing 3
9 Check weighing 4
See {6. Weighing}
in det
ail.
CALF-15 Preset tare 1
0 Disable
1 Not disabled
When disabled, preset tare can not be set nor recalled.
CALF-16 Distinguish between preset tare and tare 0
0 None (same)
1 Distinguished
When “1” is set (distinguished), "" illuminates on the
lower status display
section when “preset tare” is used and header of serial output becomes "PT"
(preset tare) and "T " (tare)
* Pressing [PRESET TARE] always shows preset value regardles
s of this
setting.
CALF-17
Serial interface; Print preset tare value with net
weight (Si F-01=3, rsF-01=3)
0
0 Not printed
1 Print
CALF-18
Push-zero and tare-clear function when powered on
and when the display is turned on.
0
0 Both push-zero and tare-clear are disabled.
1 Only push-zero is available.
2 Only tare-clear is available.
3 Both push-zero and tare-clear are available.
Push-zero is performed at stable reading. If
unst
able for 10 seconds or longer,
“
Err
St” (Error message) appears on the
display.
When reading is out of the range of
push-zero,
“Err PZr” is displayed on
the display.
(Press [ESC] to return to the normal mode.)