Instruction manual
Function #
Function name Range
Default
CALF-06 Zero Tracking Time 0.0 - 5.0 (sec.) 0.0
CALF-07 Zero Tracking Width 0 – 9 (1/2D) 0
Zero tracking compensation function will automatically bring the display back to zero when there are
small deviations. This function is set in combination with CALF-06 and CALF-07.
* Zero tracking is not performed with CALF-06=0.
* Its least input increment is 1/2D. (The width for setting of 1 is eq
uivalent to 1/2 of the minimum
graduation)
C-13