Scale User Manual
 33
2.  CC 0  is displayed. 
3. If necessary to change the weight value to check, press the   SAMPLE   key and 
change the weight value using the following keys. 
  SAMPLE  key  To select the digit blinking to change. 
  RE-ZERO  key  To set the value of the digit selected.   
  PRINT    key  To store the value and return to step 2. 
4. At step 2, press the    PRINT    key. The zero point 
is weighed and the weighed value is displayed for a 
few seconds. 
5. Place the displayed weight on the pan and press 
the 
 PRINT  key to weigh it. The weighed value is 
displayed for a few seconds. 
6.  end  appears. 
7.  glp   is displayed and calibration test report is 
output. 
8.  end  appears again. Remove the weight and 
press the  MODE  key to return to the weighing 
mode. 
 CC  0 
 0.0 
 CC 0 
The weighed data
is shown with āgā.
With nothing 
on the pan 
 2000.0 
 2000.0 
 2000.0 
Displayed 
weight 
 end 
 end 
The weighed data
is shown with āgā.
Returns to   
the weighing mode 
 glp 
GLP output