Scale User Manual
 24
Data format   
S T 
, 
+ 0 0000
.
00
g C
R
 L
F
   Header      Data      Unit Terminator 
  There are four types of headers: 
  ST : Stable weighing data (including % data) 
  QT : Stable counting data 
  US : Unstable weighing data (including count and %) 
  OL : Out of weighing range (Over) 
  The data is normally 9 digits including a decimal point and a sign. 
  There are 14 types of units:   
    g  : Weighing data “gram”   
   P C  : Counting data “pcs” 
    %  : Percentage data “%” 
    o z  : Weighing data “decimal ounce” 
   l b  : Weighing data “decimal pound” 
  o z t  : Weighing data “troy ounce” 
   c t  : Weighing data “carat” 
  m o m  : Weighing data “momme” 
  d w t  : Weighing data “penny weight” 
  G N  : Weighing data “grain” 
    N  : Force data “Newton” 
   t l  : Weighing data “tael” 
    t  : Weighing data “tola” 
  DS  : Calculated density (specific gravity) value 
  The terminator is always C
R
L
F
. 
  Example of output data: 
Weighing data “gram”  ST , +001234 .  5 
g
C
R
L
F
Counting data  Q T , + 0 0 01234 5 
P C
C
R
L
F
Percentage data  S T , + 0 0 0123 . 4 
%
C
R
L
F
Out of range “gram” (+)  O L , + 9 9 9999 . 9 
g
C
R
L
F
Out of range “pcs” (-)  O L , - 9 9 99999 9 
P C
C
R
L
F
Data output mode 
 Stream mode   
Set the function “prt 0”. 
The balance outputs the current display data. The data-update rate is approximately 
10 times per second. This rate is the same as the display-update. 
The balance does not output data while it is in the setting mode. 
Separator
(“ ” shows a space.)