User guide

9-2. Function list
Item •
Description
Auto power-off
function
F I- 0
Auto power-off disabled Automatically
power off
F 1 - 1
Auto power-off enabled for battery
use only
F /- F-
Auto power-off enabled for battery
and AC adapter
Display
resolution
F E- B
Normal (1/3,000 class)
Legal for trade
will be Normal.
*
F E- 1
High (1/6,000-1/7,500 class)
F E- E
Higher (1/12,000-1/15,000 class)
Weighing unit
when powered
on
^
F 3- Q
lb ,
U.S.A. models
only
F 3- 1
oz
F 3- E
kg
______________________________
RS-232C
Baud rate
*
F 3- B
2400 bps
F 3- 1
4800 bps
F 3- E
9600 bps
RS-232C
Data output
mode
*
F 5- B
Stream mode
F
5- /
Command mode
F 5- г
Print switch mode
F 5-
3
Auto-print mode +/- data
F 5- 3
Auto-print mode + data
Comparator
mode
*
F B- B
Comparator disabied
Conditions to
compare,
d = minimum
display division
F B- 1
Compares all data
F B- E
Compares data > +4d or < -4d
F B- 3
Compares all stable data
F B- 3
Compares stable data > +4d or < -4d
Filtering to
weighing data
F 1- B
Weak / fast response
*
F 1- 1
Normal / normal response
F E
Strong / slow response
RS-232C
Data format
^
F B- B
Reply is sent
Reaction to the
command
F B- 1
No reply except "Q” command
Factory setting