Instruction manual

59
14-1-10 Initialization
15. Initialization restores the various settings to the factory setting
values.
1> Moves to the display check mode without performing the
initialization.
1 Enters the initialization mode and displays init f in
the main display.
ESC Returns to the CHeCk display.
16. Using the 1> key, select an item to be initialized. In the example
display shown to the right, initialization of the calibration settings is
selected.
init f Initializes the function settings.
iinit C Initializes the calibration settings.
iinitda Initializes the DAV settings.*
Note
iinit0p Initializes the option settings.*
Note
iinit a Initializes all the settings.*
Note
*Note: All of the three initialization modes initialize the DAV
settings. Thus, a high-precision digital multimeter is
required.
1> Selects an item to be initialized.
1 Performs initialization. During initialization, - - - - - --
appears.
ESC Returns to the CHeCk display.
init
init f
CHeCk init
init C
CHeCk init
init C
CHeCk init
init C
CHeCk init
init C
------
init
ESC
CHeCk