40
10-4-2 Example of the 2D comparator mode (By time control)
Example: f-16 7 (5-level continuous comparison) with the starting and ending time for each level
that are set as below.
Level 1 t1 5t 60 t1 end 180 Unit: ms
Level 2 t2 5t 240 t2 end 340
Level 3 t3 5t 440 t3 end 580
Level 4 t4 5t 680 t4 end 920
Level 5 t5 5t 1100 t5 end 1340
When the judgment result is OK:
Hi 2
lo 2
Hi 1
lo 1
Hi 3
lo 3
Hi 4
lo 4
Hi 5
lo 5
COMP ON
HI output
OK output
LO output
ON
OFF
ms
ON
OFF
ON
OFF
ON
OFF
0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400