7-5 Full Calibration Mode
Zero and span calibration are carried out using an actual load after minimum division and rated
capacity have been set.
7-5-1 Setting the minimum division
1. With fCal displayed, press the key to enter the full
calibration mode.
2. When d 01 is displayed, setting the minimum division is enabled.
Use the following keys to set the minimum division.
1 Increases the value.
1 Decreases the value.
1 Saves the minimum division and goes to step 3.
ESC Cancels the operation and returns to the fCal display.
7-5-2 Setting the rated capacity
3. Use the following keys to set the rated capacity.
1> Selects the digit to set.
1 Increases the value.
1 Decreases the value.
1 Saves the rated capacity and goes to step 4.
ESC Cancels the operation and returns to the fCal display.
10000
fCal Cap
15000
fCal Cap
To set the rated capacity
Zero calibration
fCal
d 01
fCal diU
d 02
fCal diU