Instruction manual

20
7-3-3 Zero calibration
4. With nothing placed on the load cell, press the 1 key.
1 Performs zero calibration and goes to step 5.
ESC Cancels the operation and returns to the d-5p display.
7-3-4 Digital span calibration
5. Use the following keys to enter the sensor’s rated output voltage.
1> Selects the digit to set.
1 Increases the value.
1 Decreases the value.
1 Saves the rated output voltage and goes to step 6.
ESC Cancels the operation and returns to the d-5p display.
6. end is displayed. Press the 1 key to return to the d-5p
display.
Press the 1> key to go to the calibration mode.
Press the ESC key to return to the measurement mode.
Digital span calibration
------
d-5p d5pn
32000
d-5p d5pn
end
d-5p
d-5p
Cal0
d-5p