Instruction manual
19
7-3 Digital Span Mode
The sensor’s rated output voltage is keyed in. Calibration is performed without using an actual
load.
7-3-1 Setting the minimum division
1. With d-5p displayed, press the ↵ key to go to the digital span
mode.
2. When d 01 is displayed, setting the minimum division is enabled.
Use the following keys to set the minimum division.
1∧ Increases the value.
1∨ Decreases the value.
1↵ Saves the minimum division and goes to step 3.
ESC Cancels the operation and returns to the d-5p display.
7-3-2 Setting the rated capacity
3. Use the following keys to set the rated capacity.
1> Selects the digit to set.
1∧ Increases the value.
1∨ Decreases the value.
1↵ Saves the rated capacity and goes to step 4.
ESC Cancels the operation and returns to the d-5p display.
To set the rated capacity
Zero calibration
10000
d-5p Cap
15000
d-5p Cap
d-5p
d 01
d-5p diU
d 02
d-5p diU