7-2 Calibration Modes
There are three calibration modes.
Digital span mode (d-5p)
The sensor’s rated output voltage is keyed in. Calibration is performed without using an actual
load.
Calibration mode (Cal)
Zero and span calibration are carried out using an actual load.
Full calibration mode (fCal)
Zero and span calibration are carried out using an actual load after minimum division and rated
capacity have been set.
7-2-1 Selecting or changing the calibration mode
1. In the measurement mode, press the FNC key for three seconds or
more to enter the function selection mode.
2. Press the > key to display d-5p (digital span mode).
3. With d-5p displayed, press the > key to display Cal
(calibration mode).
4. With Cal displayed, press the > key to display dfCal (full
calibration mode).
FNC
4532
d-5p
Cal
fnc
fCal
3 seconds