Specifications
 80 BM series 
Display Error code  Description 
Unit mass error 
The sample unit mass for the counting mode is too light. Storing and 
using it for counting will cause a counting error.   
Add samples to reach the specified number and press the 1PRINT1 key. 
Pressing the 1PRINT1 key without adding samples will shift the balance to 
the counting mode. But, for accurate counting, be sure to add samples. 
Automatic response adjustment zero error 
The automatic response adjustment can not be performed because there 
is something on the pan. Clear the pan. Press the 1CAL1 key to return to 
the weighing mode. 
(Check NG) 
Automatic response adjustment unstable error 
The automatic response adjustment can not be performed because the 
mass value is unstable. 
Check the ambient conditions such as breeze, vibration and magnetic 
fields, also check the weighing pan.   
Press the 1CAL1 key to return to the weighing mode. 
Clock battery error 
The clock backup battery has been depleted. 
Press any key and set the time and date. The clock and calendar 
function works normally as long as the AC adapter is connected to the 
balance. If this error appears frequently, contact the local A&D dealer. 
 (Blinking) 
Memory full 
The amount of weighing data in memory has reached the maximum 
capacity. Delete data in memory to store new data. For details, refer to 
"12. Data Memory". 
 (Illuminated) 
Memory full 
The amount of calibration or calibration test data in memory has reached 
the maximum capacity (50 sets). 
The data in memory will be deleted automatically to store new data. For 
details, refer to "12. Data Memory". 
Memory type error 
Type of memory set in the function table and type of data stored are 
different. For details, refer to "12. Data Memory". 
Memory data error 
A fatal error occurs in the stored data. To use the memory function, erase 
all the old data and release the error. For details, refer to "12.1. Notes 
On Using Data Memory". 
EC,E00 
Communications error 
A protocol error occurred in communications.   
Confirm the format, baud rate and parity. 
EC,E01 
Undefined command error 
An undefined command was received.   
Confirm the command. 