Specifications
BM series  65 
Entering the density of a liquid 
  Two ways to set the density of a liquid are available in the function table, "Liquid density input (ldin): by 
entering the water temperature or by entering the density directly.   
1 Press the  MODE   key as necessary to select the density mode.   
Press and hold the   MODE   key to select to "water temperature" or "liquid density". 
In the density mode : The unit is "g (gram)". The processing indicator (  ) blinks. 
Response adjustment is off. 
If you cancel the density mode, press the  MODE  key. 
Entering the water temperature ( ldin 0 ) 
2  The water temperature currently set (unit:°C, factory setting : 25°C) is displayed. 
Use the following keys to change the value.   
 RE-ZERO (+)key
....
The key to increase the temperature by one degree.   
(0°C is displayed after 99°C) 
   MODE (-)key..........The key to decrease the temperature by one degree.   
(99°C is displayed after 0°C) 
   PRINT  key............The key to store new water temperature, display   end  and return to the 
density mode. Proceed to Step 1. 
   CAL   key...............The key to cancel the change and return to the density mode.   
Proceed to Step 1. 
The relation between the water temperature and density 
°C +0 +1 +2 +3 +4 +5 +6 +7 +8 +9 
0  0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10  0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20  0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30  0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40  0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
  g/cm
3
Entering the density directly ( ldin 1 ) 
 The density currently set (unit : g / cm
3
, factory setting : 1.0000g / cm
3
) is displayed.   
  Use the following keys to change the value. 
  The range to set the density is 0.0000g / cm
3
 to 1.9999g / cm
3
. 
   RE-ZERO1(+)key
....
The key to set the value of the digit selected. 
   RANGE1(-)key
........
The key to select the digit to change the value. 
   PRINT  key
............
The key to store the change, display  end  and return to the density mode. 
Proceed to Step 1. 
   CAL  key
...............
The key to cancel the change and return to the density mode. Proceed to Step 1. 