12.3.  Data Memory For Calibration And Calibration Test 
Features 
 Calibration data (when and how it is performed) and calibration test 
data can be stored in memory. 
  All the data in memory is available to be output at one time to a printer 
or personal computer.   
  Up to 50 data sets of the latest calibration or calibration test can be 
stored. When the memory capacity has been reached, "fUl" is 
displayed in the upper left of the display.   
Storing the calibration and calibration test data 
Note If "Clr" appears blinking in the upper left of the display, delete the data in memory. 
  Store the calibration and calibration test data as follows: 
1  Set the "Data memory (
data)" parameter to "2". 
2  Set the "GLP output (info)" parameter to "1" or "2". 
3  With the settings above, each time calibration or calibration test is performed, the 
data is stored automatically. 
Transmitting the memory data 
Note 
  Confirm that the "Serial interface (5if)" parameters are set properly.   
Refer to "10. Function Table" and "16.2. Connection To Peripheral Equipment".   
  Confirm that the "Data memory" parameter is set to "data 2". 
1  Press and hold the  PRINT  key until  out   is displayed, then release the key.  
2 Press the  PRINT  key to display  out "n"o  with "no" blinking. 
3 Press the  RE-ZERO  key to display  out "g"o  with "go" blinking. 
4 Press the  PRINT   key to transmit all memory data using the RS-232C interface. 
5 The balance displays  Clear  when all memory data is output.   
Press the   CAL   key to return to the weighing mode. 
Deleting data stored in memory 
1  Press and hold the   PRINT  key until  out   is displayed, then release the key.   
2 Press the  SAMPLE  key to display  Clear . 
3 Press the  PRINT  key to display  Clr "n"o  with "no" blinking. 
4 Press the  RE-ZERO  key to display  Clr "g"o  with "go" blinking. 
5 Press the  PRINT  key to delete all data. 
6  The balance displays  out   and returns to the weighing mode when all 
the data has been deleted. 