Specifications
BM series  57 
Title block and end block 
  When mass values are recorded as GLP data, a "Title block" is inserted at the beginning and an 
"End block" is inserted at the end of a group of mass values in the GLP report. 
Notes 
  To output the report to an AD-8121B , use MODE 3 of the AD-8121B. If MODE1 is used, select 
temporary dump print mode by pressing the  STAT. key of the AD-8121B. 
  If the data memory function is used (except data 0), the "Title block" and "End block" can not be 
output. 
Key operation 
1 With the weighing data displayed, press and hold the  PRINT   key, then release the key. 
    5tart   is displayed. The "Title block" is output. 
2  The weighing data is output according to the parameter setting of the data output mode (prt) of 
the function table.   
3  Press and hold the   PRINT  key until   recend   is displayed, then release the key. 
The "End block" is output. 
 Setting of "info 1" Setting of "info 2" 
    AD-8121 printer format General format 
    Space, ASCII 20h 
<TERM>
  Terminator, CR , LF or CR 
 CR Carriage return, ASCII 0Dh 
  LF  Line feed, ASCII 0Ah 
-----------A-&-D 
MODEL-----BM-300 
S/N-----01234567 
ID------LAB-0123 
DATE--2011/02/20 
START 
TIME----12:34:56 
WT--+123.4567--g 
WT--+123.4612--g 
WT--+123.4623--g 
WT--+123.4634--g 
WT--+123.4511--g 
WT--+123.4534--g 
WT--+123.4712--g 
WT--+123.4645--g 
END 
TIME----12:40
15 
SIGNATURE 
--------------- 
Title block
Manufacturer 
Model 
Serial number 
ID number 
Date 
Time 
Weighing data 
Time 
Signature 
End block 