Specifications
BM series  47 
MT format  5if type 3 
This format is used when the peripheral equipment of other manufacturer is connected. Please 
note that the connection is not guaranteed. 
  A header of two characters indicates the balance condition. 
  The polarity sign is used only for negative data.   
  The weighing data uses spaces in place of the leading zeros.   
  The character length of this format changes dependent upon the unit 
NU (numerical) format  5if type 4 
  This format outputs only numerical data. 
  This format consists of ten characters excluding the terminator. 
  The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign 
is used.   
CSV format  5if type 5 
  This format separates the data of A&D standard format and the unit by a comma ( , ).   
  This format outputs the unit even when the data is overloaded. 
  When a comma ( , ) is selected for the decimal point, the separators are set to semicolon ( ; ). 
  When the ID number, data number, time and date are added at "Data output (dout)" of the 
function table, outputs ID number, data number, date, time and weighing data in this order and 
separates each item by a comma and treats all the items as one group of data. 
ID number  dout 5-id 1 
  The number to identify a specific balance.   
  This format consists of eight characters excluding the terminator. 
Data number    dout d-no 1 
  This format outputs the data number just before the data is transmitted using the RS-232C 
interface. 
  This format consists of six characters excluding the terminator. 
  When CSV format (5if type 5) is selected, the period ( . ) is replaced with a comma ( , ). 
Data number  Terminator 
ID number  Data number  Date  Time  Weighing data 
Stable header 
Unstable header 
Overload header 
Header Data  Unit Terminator 
Data Terminator 