Specifications
BM series 43
Class Item and Parameter Description
off
No limit
t-Up
Timeout
on
1 second
Selects the wait time to receive
a command.
off
No output
5if
Serial
interface
erCd
AK, Error code
on
Output
AK: ASCII code 06h
0 Water temperature
d5 fnc
Density
function
ldin
Liquid density input
1 Liquid density
Available only when density
mode is selected
Unit
Unit
Refer to "5. Weighing Units".
C5 in
Internal mass correction
Displayed only when the internal mass value correction
switch is set to 1. Refer to "8. Calibration (To Adjust The
Weighing Value)".
id
ID number setting
Refer to "11. ID Number And GLP Report".
0 3 seconds
1 10 seconds
ion
Neutralization time
2 Manual operation (max. 10 min.)
Refer to "6.5. Built-in DC
Static Eliminator" ".
off
The ION key is not used.
8922
AD8922A control
on
Applies the ION key to the MODE key on AD-8922A.
0 Applies the PRINT key and RE-ZERO key.
1 Applies the PRINT key and ION key.
ionfnc
Static
eliminator
e-5
External control input
2 Applies the ION key and RE-ZERO key.
off
No used
door
Door state
function
d-dt
Door sensor
on
Used
Refer to "6.1. Basic Operation"
off
No used
locfnc
Password function
on
Used
When turning on the balance,
the password input (
pa55
)
is displayed.
loc no.
Password
Refer to "15.2. Changing Password".
:
Factory settings.
10.3. Description Of The Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target mass, weighing a very light sample or when quick response weighing is
required. After setting, the balance displays FAST .
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts. After setting, the balance displays SLOW .
Notes In automatic response adjustment, this parameter is selected automatically.