Specifications
BM series  31 
8. Calibration (To Adjust The Weighing Value) 
8.1.  Calibration Group 
Calibration 
 Automatic self calibration (Calibration due to changes in temperature)   
 Calibration using the internal mass (One-touch calibration) 
 Calibration using an external weight that you have 
Calibration test 
 Calibration test using the internal mass   
  Calibration test using target mass that you have 
Correction of the internal mass value 
  Correction of the internal mass value 
Caution 
  Do not allow vibration or drafts to affect the balance during calibration. 
 Calibration test does not perform calibration. 
  When using the data output for GLP / GMP using the RS-232C interface, set "GLP output (info)" 
of the function table. Refer to "10. Function Table". Time and date can be added to the GLP 
report. If it is incorrect, refer to "10.7 Clock And Calendar Function" and adjust them. 
 Calibration test is available only when "GLP output (info)" is set . 
  The calibration and calibration test data can be stored in memory. When using memory, set "Data 
memory (data)" of the function table. Refer to "12. Data Memory" for details. 
Caution on using an external weight 
  The accuracy of an external weight can influence the accuracy of weighing. 
  Select a mass for calibration and calibration test from the following table. 
Model Usable calibration mass Adjustable range 
BM-20    1 g  2 g  5 g  10 g  20 g
*
BM-22    1 g  2 g  5 g  10 g  20 g
*
  -3.000 mg  to  +3.099 mg 
BM-252    10 g  20 g  50 g  100 g  200 g
*
    -15.00 mg  to  +15.99 mg 
BM-200    50 g  100 g  200 g
*
BM-300    50 g  100 g  200 g
*
 300 g 
BM-500    50 g  100 g  200 g
*
  300 g  500 g 
  -30.0 mg  to  +30.9 mg 
  Bold type*: Factory setting. 
    The calibration mass value can be adjusted within the range above. 
About the internal mass 
  The internal mass may change due to corrosion or other damage caused by the operating 
environment, or due to aging. Check the internal mass periodically and correct the internal mass 
value if necessary. Refer to "8.7. Correcting The Internal Mass Value". 
Display 
 This indicator means "In process of measuring calibration data". 
Do not allow vibration or drafts to affect the balance while the indicator is 
displayed. 