Specifications
BM series 9
Step
3 If the separation plate is not necessary, remove
the screw and take it off. Refer to next page
"The Separation Plate".
Step
4 Adjust the level of the balance using the leveling feet.
Step
5 Confirm that the AC adaptor type is correct for your
local voltage and power receptacle type.
Step
6 Connect the specified AC adapter to the balance.
Warm up the balance for at least one hour with
nothing on the weighing pan.
Step
7 Confirm correct weighing. If necessary, calibrate
the balance. Refer to "8. Calibration".
Screw
Hoo
k
Separation plate
Bubble spirit level
Leveling foo
t
A
C adaptor jac
k
Grounding terminal
Example of outle
t
Note Please confirm that the AC adapter
type is correct for your local voltage
and receptacle type.
BM-200, BM-300,
BM-252
BM-500, BM-252
BM-22, BM-20