Product specifications

Page 86
Command Protocol
RTTL
Read total data of code
138 characters.
This command outputs total data of a selected code.
Parameters are 7 figures with "-" sign at MSB when negative
value and without decimal point. If code number is replaced with
space codes (20h), the current total data of code are output.
Code number, Total count, Total weight
RERR
Read error code
12 characters.
This command outputs current error code number and type.
Each left figure 0: normal, 1: error
Each right figure is error code.
Alarm2, Alarm1, Zero error, Weighing sequence error
Example: Above code indicates weighing sequence error and
"Weighing sequence is stopped" of error code 0.
RERR
C
R
L
F
RERR 00 00 00 10
C
R
L
F
Computer
AD-4404
RTTL0034
C
R
L
F
RTTL0034,1234567,...,123456789
C
R
L
F
Computer
AD-4404
Output item order Length
Code number RCOD + 4
Total count 7
OK count 7
NG count 7
Hi count 7
Lo count 7
HiHi count 7
LoLo count 7
Foreign matter count 7
Duplication count 7
Crush count 7
Maximum 7
Minimum 7
Average 7
Standard deviation σ
n-1
7
Population standard deviation σ
n
7
Total weight 9