Product specifications

Page 79
9.2.5.
9.2.5.9.2.5.
9.2.5. General Data Format
General Data FormatGeneral Data Format
General Data Format
This format is used for the command mode and jet stream mode.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28
R G R S 0 0 9 9 , 1 2 3 4 5 6 7 , 1 2 3 4 5 6 7 8 9
C
R
L
F
Header
Command is echoed. The echoed command is 4 characters
Code number
The code is 4 characters. A comma "," is appended after this.
Weighing data
Data uses BCD code, is 7 figures and does not include a decimal point.
When data is negative, a minus sign is appended to the head.
A comma "," is appended after this.
Status
The status of weighing data and sequence are indicated at 36 bits.
The status bits synchronizes with control I/O and options.
9 figure is ASCII code. Each lower 4 bits are used.
Each upper 3 bits are fixed to "101".
Terminator
Select a terminator at Terminator [r5 f-7]
CR or CR + LF CR: 0Dh, LF: 0Ah
Header Code number Weighing data
System Monitor
Terminator
2
0
: Stable
2
1
: Zero band
2
2
: Full
2
3
: LoLo
1 2 8 9 7 6 5 4 3
2
0
:Lo
2
1
: OK
2
2
: Hi
2
3
: HiHi
2
0
: In operation
2
1
: Conveyor
2
2
: Busy
2
3
:
2
0
: Foreign material
2
1
: Crush
2
2
: NG
2
3
: Target count
2
0
:
2
1
:
2
2
: Online
2
3
: In weighing sequence
2
0
: Internal reservation
2
1
: Weighing error
2
2
: Alarm 1
2
3
: Alarm 2
2
0
: Zero error
2
1
: Out of range
2
2
: Buzzer
2
3
: Tare entered
2
0
: Center of zero
2
1
: Gross display
2
2
: Net display
2
3
: In hold display
Internal reservation
Example: 2
3
2
2
2
1
2
0
1
0
1
0 = Ah
ASCII code = 3Ah.
Upper 4 bits are 3h always.