Product specifications

Page 63
8.17.
8.17.8.17.
8.17. Tare Operation
Tare OperationTare Operation
Tare Operation
The relation of the display is as follows:
Net = Gross - Tare
Key Operation
Press the TARE key.
Operation from Control I/O
[1n f-xx]
[03]
In xx (Axx) .................The input terminal of the control I/O.
Tare ................................When it turns on, zero of net is displayed.
xx is terminal number (01 to 11).
[Function] - [Set function] - [control I/O] - [Input]
Concerning Parameters
[genf-9]
Tare & zero unstable WGT
The prohibition of the re-zero operation in the unstable condition.
[Function] - [Set function] - [General] - [Weight]
[genf-10]
Tare at (-)gross weight
The prohibition of the tare operation during negative weighing.
[Function] - [Set function] - [General] - [Weight]
[genf-13]
Clear mode at power ON
Specify power-on status.
[Function] - [Set function] - [General] - [Weight]
8.17.1.
8.17.1.8.17.1.
8.17.1. Tare Clear Operation
Tare Clear OperationTare Clear Operation
Tare Clear Operation
The function displays a gross value that is not compensated by the tare operation or
the zero operation.
Key Operation
Step 1 Turn off the display.
Step 2 Pressing and holding the TARE key, press the ON key.
Operation from Control I/O
[1n f-xx]
[04]
In xx (Axx) .................The input terminal of the control I/O.
Tare clear .................When it turns on, the original gross is displayed.
xx is terminal number (01 to 11).
[Function] - [Set function] - [control I/O] - [Input]