8.16.
8.16.8.16.
8.16. Zero Operation
Zero OperationZero Operation
Zero Operation
The zero operation zeroes the display and changes to the gross display.
The operation can be performed from the front panel key or the input terminal of the
control I/O.
The adjustable range is based on the zero calibration and Zero range [genf- 6].
The range is displayed in the unit of percentage of the weighing capacity.
If the operation can not be achieved, an error message and ZR.ERR is displayed.
The re-zero data is maintained in memory even without power.
Key Operation
Pressing and holding the SHIFT key, press the ZERO key.
Operation from Control I/O
[1n f-xx]
[01]
In xx (Axx) .................The input terminal of the control I/O.
Zero ................................When it turns on, zero of gross is displayed.
xx is terminal number (01 to 11).
[Function] - [Set function] - [control I/O] - [Input]
Concerning Parameters
[genf-1]
Zero range
Range is within 30% of the weighing capacity.
[Function] - [Set function] - [General] - [Weight]
[genf-9]
Tare & zero unstable WGT
The prohibition of the re-zero operation in the unstable condition.
[Function] - [Set function] - [General] - [Weight]
[genf-13]
Clear mode at power ON
Specify power-on status.
[Function] - [Set function] - [General] - [Weight]