Product specifications
Page 48
7.3.
7.3.7.3.
7.3. Recalling a Code
Recalling a CodeRecalling a Code
Recalling a Code
The method to recall a code can be selected at Code recall method [5qf-81].
Function and parameter
5qf-81
1
Code recall method
Key/Serial I/F
5qf-81
2
Code recall method
Parallel I/F
5qf-81
3
Code recall method
External switch control
[Function] - [Set function] - [Sequence] - [Other]
Detail of Parameters
[5q f-81]
[1]
Code recall method
Key/Serial I/F
A code is recalled with key operation or input data of the interface.
The last data input has effect.
In case of using the Key Operation:
Press the CODE RECALL. Enter a code number and press ENTER key.
In case of using the Serial Interface:
The code number can be entered by command mode.
Built-in RS-485.
RS-422/485 interface of OP-03.
RS-232C interface of OP-04.
In Case of Field Bus Interface:
The code number is specified by PLC data.
CC-link of OP-20.
DeviceNet of OP-21.
PROFIBUS of OP-22.
[5q f-81]
[2]
Code recall method
Parallel I/F
A code is recalled with input data of the control I/O or parallel interface.
Use BCD code.
In Case of Parallel Interface:
The code number can be specified with BCD code. Example: encoder switch.
Control I/O.
Parallel interface of OP-05.
[5q f-81]
[3]
Code recall method
External switch control
Method of 1 or 2 can be selected at an input terminal specified to code
number 48 in control I/O or OP-05. Terminal status is as follows:
Off: 1, Key/Serial I/F
On: 2, Parallel I/F. Code number of 2 is kept until next input.