Product specifications

Page 144 AD-4404
Category address
:
[Function] - [Set Function] - [Serial] - [RS-485]
Category
address
symbol
Name Descriptions
Range
and
choices
Default
r5 f- 1 Output data
When jet stream mode of [r5 f- 2] is
used, 1, 2 or 3 can be selected. When
the display value is held in jet stream
mode, output is not stopped.
1: Displayed value
2: Gross value
3: Net value
4: Tare value
5: Gross value/ Net value/ Tare value
6: Displayed value with code
7: Gross value with code
8: Net value with code
9: Tare value with code
10: Gross value/ Net value/ Tare value
with code
1 to 10 1
r5 f- 2
Communication
mode
1: Stream mode
2: Auto print mode
3: Manual print mode
4: Jet stream mode
5: Command mode
6: MODBUS
1 to 6 6
r5 f- 3 Baud rate
1: 600 bps
2: 1200 bps
3: 2400 bps
4: 4800 bps
5: 9600 bps
6: 19200 bps
7: 38400 bps (Jet stream mode only)
1 to 7 5
r5 f- 4 Parity check
0: Not used
1: Odd
2: Even
0 to 2 0
r5 f- 5
Character
length
7: 7 bits
8: 8 bits
7, 8 8
r5 f- 6 Stop bits
1: 1 bit
2: 2:bits
1 to 2 1
r5 f- 7 Terminator
1: CR CR: 0Dh
2: CR LF LF: 0Ah
1 to 2 2
r5 f- 8 Address
0: Address not used
1 to 99: Address used
0 to 99 0
r5 f- 9
Response
timer
Set the waiting timer from receiving
command to transmitting a response.
0.0 to
25.5 s
0.0 s
r5 f- 10
r5 f- 11
Do not change. Internal reservation.