Product specifications

Page 124 AD-4404
11.
11.11.
11. Function List
Function ListFunction List
Function List
The function list stores parameters to control the indicator.
The parameters are stored in an item even without power supplied.
An item is classified by a category address, and is further classified by an item
number. Refer to " 11.1.2. Outline of the Function List".
The category address has a symbol for the 7-segments display.
There are two kind of the function modes to operate the function list.
Parameter settings The mode to edit the parameter.
Referring parameters The mode to refer to the parameter during the sequence.
Example of the Display Form for an Item:
Category address:
[Function] - [Set function] - [General] - [Weight]
genf- 1
Display refresh rate
5 times / sec
1
Caution
When entering Set function of the function mode, the current weighing
sequence is stopped.
11.1.1.
11.1.1.11.1.1.
11.1.1. Operation Keys
Operation KeysOperation Keys
Operation Keys
To enter the function list Press and hold the ENTER key and press the
key in the weighing mode.
Select the menu Function using the key and
the ENTER key.
To select the parameter address The , SHIFT + , ENTER, ESC keys.
To change the parameter The , SHIFT + , Alphanumerical, ENTER,
ESC keys.
To store it and exit the mode The ESC key.
(To return to weighing mode)
Item number
Category address symbol
Meaning of the Item number
An Item
Parameter
Address of the Item
Meaning of the parameter