Product specifications
AD-4404 Page 123
10.5.
10.5.10.5.
10.5. Remote Operation
Remote OperationRemote Operation
Remote Operation
This mode can read and write the parameters of the function list, the code data and
calibration data.
The built-in RS-485, RS-422/485 (OP-03) or RS-232C (OP-04) is used for remote
operation.
It is necessary to install the remote setup program in the computer or controller
before use. Refer to http://www.aandd.co.jp
Refer to the instruction manual for details of the program.
Caution
Do not download data during a weighing operation.
Do not turn off the indicator during down load.
Remove all connections to prevent an irregular operation.
Maintain the power supply during the remote setup operation.
Entering the Remote Operation Mode
Step 1 Press and hold the ENTER key and press the key to display the
menu in a weighing mode.
Step 2 Select the menu SystemMgmt using the key, ENTER key.
Select the menu RemoteSetup using the key, ENTER key.
Category address:
[Function] - [Set function] - [SystemMgmt] - [RemoteSetup]
Step 3 Select a menu using the key, ENTER key.
Menu: Data or Software
Step 4 If Data is selected, press Yes to down load the data using the
key, ENTER key.
Step 5 Press the ESC key to return to weighing mode.
Advise
The following RS-232C to RS-485 converter can be used.