Product specifications
Page 122 AD-4404
Procedure
Caution
Do not initialize parameters while in operation. Cut off the power supply of
other systems. When initializing the indicator, the output may change.
When initializing the indicator, do not turn off the indicator before it is reset.
To enter initialization
Step 1 Press and hold the ENTER key and press the key to display the
menu in a weighing mode.
Step 2 Select the menu SystemMgmt using the key, ENTER key.
Select the menu Init using the key, ENTER key.
Category address:
[Function] - [Set function] - [SystemMgmt] - [init]
Step 3 Select the initialization menu using the key, ENTER key.
Menu: RAM, Code, Function, CAL or All
Step 4 Press Yes to initialize them using the key, ENTER key.
Step 5 Wait for the indicator to reset.