Product specifications

AD-4404 Page 119
10.3.4.
10.3.4.10.3.4.
10.3.4. Testing A/D Converter
Testing A/D ConverterTesting A/D Converter
Testing A/D Converter
The A/D converter data is displayed.
When pressing the ENTER key, a test voltage can be input to the A/D converter.
10.3.5.
10.3.5.10.3.5.
10.3.5. Testing BCD Output of OP
Testing BCD Output of OPTesting BCD Output of OP
Testing BCD Output of OP-
--
-01
0101
01
One of terminal A and B alternately turns on every one second and in order. The
active terminal displays "1".
10.3.6.
10.3.6.10.3.6.
10.3.6. Testing Relay Output of OP
Testing Relay Output of OPTesting Relay Output of OP
Testing Relay Output of OP-
--
-02
0202
02
A terminal of the relays turns on every one second and in order. The active terminal
displays "1".
10.3.7.
10.3.7.10.3.7.
10.3.7. Testing RS
Testing RSTesting RS
Testing RS-
--
-422/485 Interface of OP
422/485 Interface of OP422/485 Interface of OP
422/485 Interface of OP-
--
-03
0303
03
Each time to press the ENTER key, the test data "ST,GS,+0000000kg CR LF" is
output.
The received data is displayed, when data is output.
: CR
: LF
16.001kg
A/D Count L/C Output(mV/V)
964583 3.103872
Int.Count Test Input
160008 0ff
A/D converter output
Internal count
Loadcell output
Test input
12345kg
OP-01 000000000000000000
a18 a1
000001000000000000
b18 b1
A1
B1
A18
B18
123kg
OP-02 0001 00000
10 7 5 1
Output terminal
1 ~ 5
Output terminal
7 ~ 10
Sent data
123kg
RNET
ST,GS,+0000000kg
0p-03 rcu and 5nd
Received data
Parity: p
Framing:
f