AD-4212C-300 AD-4212C-600 AD-4212C-3000 AD-4212C-6000 AD-4212C-301 AD-4212C-3100 Production Weighing Unit INSTRUCTION MANUAL 1WMPD4001992B
© 2011 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company, Limited. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice. Windows and Excel are registered trademarks of Microsoft Corporation. .
CONTENTS 1. INTRODUCTION ..........................................................................................................................2 1.1. Features ........................................................................................................................................ 2 1.2. Compliance ................................................................................................................................... 5 2. UNPACKING THE BALANCE .............................
1. INTRODUCTION This manual describes how the AD-4212C series balance works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the balance and keep it at hand for future reference. 1.1. Features z A Weighing Unit, suitable for building into a production line system. The weighing unit is compact, with a width of 59 mm. z A Super Hybrid Sensor (SHS) is used to provide high resolution and high response speed.
Example 1 (Connecting to a personal computer or an AD-8922A) AD-4212C Accessory RS-232C cable (10 m) Communication software “ WinCT-AD4212C” USB converter (Option: AX-USB-9P) USB port Personal computer AD-8922A AD-4212C USB converter (Option: AX-USB-9P) Accessory RS-232C cable (10 m) Communications cable (Option: AX-KO1786-200) 3 Personal computer
Example 2 (Connecting to a PLC) . D-sub 9-pin (female) AD-4212C Accessory RS-232C cable (10 m) PLC AD-8922A-01 Connection cable* AD-4212C BCD input AD-8922A PLC BCD output board (Option AD-8922A-01) BCD cable (Fabricate locally) * Specify the AD-8922A-01 connection cable when ordering the AD-8922A-01.
1.2. Compliance Compliance with FCC Rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when the equipment is operated in a commercial environment.
2. UNPACKING THE BALANCE 2.1. Unpacking The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used for transporting the balance in the future. When unpacking, see the illustration below to confirm that everything is contained. AC adapter jack Weighing pan Screws (Flat head: M4 X 6) Pan support RS-232C connector AC adapter plug Grounding terminal (See note below.
ATTACHING THE AD-4212C SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE When the AD-4212C series weighing unit is built into a system, remove the three leveling feet and use the screw holes to secure the weighing unit to the base. Attachment Procedure 1. Remove the three leveling feet. Caution Do not remove any other screws from the bottom of the weighing unit. Doing so could damage the weighing sensor. Weighing unit 2. Secure the weighing unit from above.
2.2. Installing the Balance Install the balance as follows: 1. Refer to “3. PRECAUTIONS” for installing the balance. 2. Refer to "2-1 Unpacking" on page 6, to attach the pan support and the weighing pan on the weighing unit. 3. Insert the RS-232C cable into the jack located on the rear of the weighing unit. Rear of the weighing unit When confirming the performance of the weighing unit using WinCT-AD4212C, connect the cable into RS-232C terminal of the computer.
MASS OF THE LOCALLY FABRICATED WEIGHING PAN Design the weighing pan so that the mass falls within the ranges shown in the table below: Model Mass of the weighing pan (g) 20 to 130 Status Pan support is used (with weighing pan removed) Pan support boss is used (with the weighing pan and pan support removed) Pan support is used (with weighing pan removed) AD-4212C-600 Pan support boss is used (with the weighing pan and pan support removed) Pan support is used (with weighing pan removed) AD-4212C-3000/3100
3. PRECAUTIONS To get the optimum performance from the balance and acquire accurate weighing data, note the following: 3.1. Before Use Install the weighing unit in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity. Install the weighing unit where it is not exposed to direct sunlight and it is not affected by heaters or air conditioners. Install the weighing unit where it is free of dust.
3.2. When Building Into a System The AD-4212C is a precision balance. When it is built into a system, errors such as unstable weight values may occur due to static electricity, vibration, sample temperature and magnetism of the materials weighed or used for the devices near the balance.
Errors due to a static charge and measures to take When the ambient humidity is less than 45% RH, insulators such as plastic or glass are prone to static electricity. When charged material comes close to the balance, a pulling force is generated by static induction between the charged material and the weighing pan. This causes an unstable weight value.
Errors due to air flow and measures to take Where the influence of ambient air flow is great such as: close to an air conditioner, door or passage way. Even very subtle air flow that is hard to be detected may influence the weighing operation. Avoid those areas as a weighing site. If weighing is to be performed in such an area, use a breeze break or take other appropriate measures. Where the influence of heat or drafts is great Eliminate temperature differences Draft between a sample and the environment.
3.3. During Use To minimize the affect by electrical noises, earth ground the weighing unit and the display unit (option). Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the balance weighing capacity. Place a sample in the center of the weighing pan to minimize corner-load errors. To prevent possible errors, before each weighing, perform re-zero using the RS-232C command, or calculate the difference between the weight value before and after weighing.
4. OPERATION OF WinCT-AD4212C The software has the following functions: - Set the weighing speed - Set the minimum display value - Calibration - Set the baud rate • The parameters set using these functions, are stored in non-volatile memory, and maintained, even if the AC adapter is removed,. - A data logging function, allows confirmation of the weighing data after weighing Before use 1. Install the WinCT-AD4212C into the computer. For details, refer to Readme of WinCT-AD4212C. 2.
4.1. Changing the Weighing Speed Using WinCT-AD4212C The weighing speed can be selected from the following three rates to minimize the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed. Indicator FAST MID. SLOW Speed Fast Slow Weighing speed setting Stability Sensitive value Stable value Operation MODE 1. Press the MODE button. 2. Press the MODE button to select a weighing speed. Either FAST, MID. or SLOW can be selected.
Calibration procedure This function calibrates the balance using the calibration weight. (Display example: AD-4212C-300) Operation 1. Plug in the AC adapter and warm up the balance for 30 minutes or more with nothing on the pan. 2. Press the CAL button. Cal 0 is displayed. Press CAL If you want to cancel calibration, press the CAL button. The balance will return to the weighing mode. If you want to change the calibration mass value, press the SAMPLE button.
4.3. Changing the Baud Rate Using WinCT-AD4212C Operation 1. Click the [Settings] tab. 2. Select the baud rate to be used and click the Set [Settings] tab Set button 3. “Completed” appears when the setting is complete. 18 button.
5. SMART RANGE FUNCTION The AD-4212C-301/3100 are equipped with two ranges. The precision range has a higher resolution. The standard range has a normal resolution. The range switches automatically, depending on the value displayed. Sending the R command allows weighing in the precision range, regardless of the tare value. The range can be fixed to the standard range by changing the minimum weighing value using the SMP command.
6. COMMUNICATION SPECIFICATIONS The AD-4212C series can communicate interactively using RS-232C. The AD-4212C series can continuously transmit the weighing data. (Approx.10 times/second* to 50 times/second) The AD-4212C series is DCE. The AD-4212C series can connect directly to the RS-232C terminal of a personal computer by using the accessory RS-232C cable. * Factory setting 1.
2. Pin position Pin position of the RS-232C cable (D-Sub 9-pin side, female) 5 1 Inch screw 9 6 Pin No. Signal name Direction 1 (Vs) - 2 TXD Output 3 RXD Input 4 5 6 7 8 SG RTS CTS Input Output Description Internally used* (Power supply GND terminal for external equipment) Transmit data Receive data N.C.
4. Command The balance can be controlled by the following commands. When transmitting, add the terminator to each command. Commands to query weighing data When the display is turned on, the AD-4212C series outputs the weighing data continuously. To use the Q or S command, send the C command to stop the continuous data output, then request the weighing data using the Q or S command. Command C Q S SIR Description Cancels the SIR command. Requests the weighing data immediately.
CAL command example External devices Time Weighing unit (PC, PLC) CAL command C A L CR LF S T , + 0 0 @ R E ; 0 0 @ D P ; 0 0 0 @ S T 0 0 ; 0 0 .
SMP command example Personal computer Time Weighing unit S T , + SMP command S M P CR LF S T , 0 0 + 0 0 0 0 0 . 0 1 2 g . 0 1 g 0 0 0 g 4 g 0 0 R command example when the precision range is used Personal computer Time Weighing unit S T , + 0 0 0 . 0 The weight value exceeds 51 g. S T , + R command 0 0 5 1 . 2 3 R CR LF (While waiting for the weight value to become stable, stops the data output) S T , + 0 0 0 . 0 0 0 0 g 1 . 2 3 4 5 g Place a sample.
7. MAINTENANCE Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair. Use the original packing material for transportation. Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.
8. TROUBLESHOOTING 8.1. Checking the Balance Performance and Environment The balance is a precision instrument. When the operating environment or the operating method is inadequate, performing accurate weighing will not be possible. If the balance seems to have a problem with repeatability or to perform improperly, check as described below. Also, visit our website, http://www.aadd.co.jp/, for “Information Library” and “FAQ”. If improper performance persists, contact the local A&D dealer for repair.
Sample and container Has the sample absorbed or lost moisture? Has the temperature of the sample, and the container if used, been allowed to equalize to the ambient temperature? Refer to “3-2 When Building into a System”. Is the sample charged with static electricity? Refer to “3-2 When Building into a System”. Is the sample of magnetic material such as iron? Use much care when weighing magnetic materials. Refer to “3-2 When Building into a System”. 8.2.
9. SPECIFICATIONS AD-4212C-300 AD-4212C-600 Weighing capacity 320 g 620 g Maximum display 320.084 g 620.084 g 320.084 g / 51.0009 g Minimum weighing value (1 digit) 0.001 g 0.001 g 0.001 g / 0.0001 g Repeatability (Standard deviation) 0.001 g 0.001 g 0.001 g / 0.0002 g ±0.002 g ±0.005 g Linearity Stabilization time in seconds (typical at FAST under good environment) Display refresh rate I/O unit RS-232C Sensitivity drift AD-4212C-301 *1 320 g / 51 g 0-30 g 0.5 s 0-30 g 0.
AD-4212C-3100 AD-4212C-3000 AD-4212C-6000 Weighing capacity 3200 g 6200 g Maximum display 3200.84 g 6200.84 g 3200.84 g / 510.009 g Minimum weighing value (1 digit) 0.01 g 0.01 g 0.01 g / 0.001 g Repeatability (Standard deviation) 0.01 g 0.01 g 0.01 g / 0.002 g ±0.02 g ±0.04 g Linearity Stabilization time in seconds (typical at FAST under good environment) Display refresh rate I/O unit RS-232C 0-300 g 0.5 s 0-300 g *1 3200 g / 510 g 0.5 s 300-3200 g 1.0 s 300-6200 g 1.0 s ±0.
10. EXTERNAL DIMENSIONS Pan support dimensions 91.9 1 50 4-M4 80.5 30 59 50 46 0.2 3-M6 When securing, use the M5 screws. 15.6 66 9.3 1 50 231 0.2 Size of when connected. Approx.30 16 254 218 231 50 46 13 (6.5) 98 48.8 30 18.8 48.8 2- 4.
11. OPTIONS AX-KO3590-200 RS-232C cable 2m AX-KO3590-500 RS-232C cable 5m AX-KO3590-1000 RS-232C cable 10 m (provided with the balance as a standard accessory.) AX-USB-9P: USB converter/Cable set USB converter (D-Sub 9 pin - USB) RS-232C cable (D-Sub 9 pin - D-Sub 9-pin) * This RS-232C cable is not used with the AD-4212C series. AD-8922A: Remote controller The AD-8922A can receive the weighing data, and performs re-zero and calibration manually.
AD-1684: Electrostatic field meter This option measures the amount of the static charge on the sample, tare or peripheral equipment and displays the result. If those are found to be charged, discharge them using the AD-1683 DC static eliminator.
12. TERMS/INDEX Terms Stable value Environment The weight value when the stabilization indicator appears. Ambient conditions such as vibration, drafts, temperature, static electricity or magnetic fields which affect the weighing operation. Calibration Adjustment of the balance using the calibration weight so that it can weigh accurately. Output To output the weighing data using the RS-232C interface. Zero point A weighing reference point or the zero display.
I AX- KO3590-500 ..................................... 4, 31 AX- KO3590-1000 ................................... 4, 31 I/O unit ...................................................28, 29 AX-USB-9P.............................................. 3, 31 Insulators .....................................................12 B L LAN................................................................4 BCD..................................................... 2, 4, 31 Leveling foot ....................................
Remote controller ........................................ 31 U Repeatability.............................. 26, 28, 29, 33 U command .................................................22 RE-ZERO button ......................................... 15 Unit ..............................................................15 RS-232C.............................................. 3, 4, 20 USB ...............................................................3 RS-232C cable ....................................