Instruction manual

7
When the external 24-VDC power supply is used
Connect an external 24-VDC power supply to the DC input terminal located
on the rear of the AD-8923-CC.
Precautions on using the external power supply
CAUTION
z Use a power supply within the rated voltage range (24 VDC±10%).
Never use a power supply with a voltage exceeding the rated range.
It may cause damage or heat buildup.
The AD-8923-CC may not function properly.
z Ground the FG terminal of the switching power supply used.
To avoid electrical shock and increase the system safety.
To increase the resistance against noises.
z Do not share the power line with other devices.
Strong noises introduced from other devices may cause damage to the
AD-8923-CC.
Inrush current from other devices may cause the AD-8923-CC not to start
up properly.
Circuit configuration of the AD-8923-CC may affect other devices to
prevent them from functioning properly.
z Select a switching power supply with a capacity of approximately 700mA
for each AD-8923-CC. Note that the AD-8923-CC may not start up with a
capacity less than 700mA.
If the power supply capacity is not sufficient, the AD-8923-CC may not
function properly.
z Be sure to add a noise filter on the front end of the switching power supply
and ground the FG terminal.
This will increase the resistance against noises.
z Be sure to ground the FG terminal of the AD-8923-CC and the AD-4212C.
This will increase the resistance against noises.