7. TROUBLESHOOTING
Symptom Description
eerror 10 appears.
Communication settings of the AD-8923-CC do not
match with those of the weighing instrument.
Check the settings such as baud rate and parity
and correct them as necessary.
For details, refer to “3.1. Setting the AD-4212C and
the AD-8923-CC”.
- - - - - - (Bar display)
remains and the
weight value is not
displayed.
Is the data output mode of the weighing
instrument set to “stream mode”?
In a mode other than “stream mode”, the
weight values are displayed only when they
are transmitted.
Check if the communication settings are correct.
Check if the cables are the correct type and are
not damaged.
The display flickers.
Electrical noise may cause this symptom.
Ground the FG terminal located on the rear of the
AD-8923-CC.