Instruction manual
15
6. CC-Link CONNECTOR
The AD-8923-CC CC-Link is a remote device station of CC-Link ver.1.10.
When a CC-Link is used, the AD-8923-CC can be controlled by the PLC
remote I/O or remote registers. So, the program can be simple. And
connection to a PLC is simple, thus, a weighing system can be built easily.
The setting values of CC-Link are changed in the function setting CCl.
6.1. CC-Link interface specifications
Station number 1 to 64
Baud rate 156 kbps, 625 kbps, 2.5 Mbps, 5 Mbps, 10 Mbps
Communications connector
The connector used can be attached or removed while the power is ON.
The function of each signal line is as follows.
DA Signal DA
DB Signal DB
DG Signal ground
SLD Shield
FG Frame ground
Status LEDs
LED ON OFF Blinking
RUN Normal
• Resetting
• No signal
−
SD Transmitting − −
RD Receiving − −
ERR
• Setting error
• CRC error
• Station trouble
Normal
When the setting
values are changed
CC-Link connector and LEDs