Instruction manual
17
7. TROUBLESHOOTING
Symptom Description
eerror 10 appears.
Communication settings of the AD-8923-BCD do not match
with those of the weighing instrument.
Check the settings such as baud rate and parity and correct
them as necessary.
For details, refer to “3.1. Setting the weighing instrument and
the AD-8923-BCD”.
- - - - - - (Bar display)
remains and the weight
value is not displayed.
• Is the data output mode of the weighing instrument set to
“stream mode”?
In a mode other than “stream mode”, the weight values
are displayed only when they are transmitted.
• Check if the communication settings are correct.
• Check if the cables are the correct type and are not
damaged.
The display flickers.
Electrical noise may cause this symptom.
Ground the FG terminal located on the rear of the
AD-8923-BCD.