Instruction manual
14
6. BCD OUTPUT CONNECTOR
Outputs the weighing data received from the weighing instrument in BCD
format, along with the polarity (+/-) and the data status (stable/unstable and
over (E display)).
Using the STROBE signal, the data can be read easily. BUSY input enables
the data to be held or prevents data refreshing during the reading operation.
Contact inputs are RE-ZERO and ON/OFF. They have same function as the
key switches on the front panel.
6.1. Connector Pin No and Specifications
I/O connector of the rear panel
Pin assignments and I/O logic
Output pin assignments Input pin assignments
Pin No. Signal Pin No. Signal
26 1 7 BUSY
27 2 9 RE-ZERO
28 4 5 ON/OFF
Switch
29 8
10
0
3 Input signal GND
39 1 11 Do not use
40 2
41 4
42 8
10
1
-The pins, which are not specified, have no connection.
(2, 4, 6, 8, 10, 36, 38 pin)
12 1 Output logic
13 2 Output logic (Factory settings)
14 4 Data 1 ON
15 8
10
2
Polarity Positive or zero ON
16 1 Stability Stabilization indicator ON ON
17 2 Over e, -e ON
18 4 Strobe Changing data ON*
19 8
10
3
Status Weighing state ON
20 1
21 2
22 4
23 8
10
4
46 1
47 2
48 4
49 8
10
5
24 1
25 2
30 4
31 8
10
6
32 1
33 2
34 4
35 8
10
7
Data
50 Polarity
45 Stability
44 Over
Status
43 Strobe
Controlling
signals
1 Output signal GND
Housing Frame ground
* When changing OFF→ON, replaces the data.
- All output, open collector; withstand voltage 30 V; no
pull-up resistor; low-level output current 10 mA
Half pitch 50 pin
50
26
125
Balance
interior
0
7
Output pins
BCD
signals
Pin 1
BCD signals
(Data from 10 -1 to 10 -8, Polarity,
Stability, OVER, Status, Strobe)
Output signal GND
Use a shielded cable.
Housing (Cable shield)
FG