Instruction manual
83
AD-4212B-23 AD-4212B-101 AD-4212B-102 AD-4212B-201 AD-4212B-301
Weighing capacity
21 g 110 g / 31 g*1 110 g 210 g 310 g
Maximum display
21.000084 g
110.0084 g /
31.00009 g*1
110.00084 g 210.0084 g 310.0084 g
Minimum weighing value
(1 digit)
0.001 mg
0.1 mg /
0.01 mg*1
0.01 mg 0.1 mg
30 g 0.02 mg
Repeatability
(Standard deviation)
Loaded on the pan
0 to 10 g:
0.003 mg*2
10 g or greater:
0.004 mg*2
0.1 mg /
0.05 mg*1
100 g 0.04 mg
0.1 mg 0.2 mg
Linearity
±0.01 mg
±0.2 mg /
±0.05 mg*1
±0.1 mg ±0.2 mg ±0.3 mg
Stabilization time (typical at
FAST under good
environment)
12 seconds
2.5 seconds /
4.0 seconds*1
4.0 seconds 2.5 seconds
Display refresh rate
10 times/second (5 times/second can be selected)
RS-232C
Bi-directional, 600-19200 bps
Comparator contact output
3-level or 5-level output
I/O
unit
External contact input
RE-ZERO, PRINT
Buzzer Sounds the connected buzzer for 3-level or 5-level comparison results
Sensitivity drift
±2 ppm/°C (10°C-30°C/50°F-86°F )
Operating environment
5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation)
Calibration weight provided
(Conforming to OIML Class E2)
1 g 50 g 100 g
Applicable weight values
1 g, 2 g, 5 g,
10 g, 20 g
10 g, 20 g, 50 g, 100 g
10 g, 20 g,
50 g, 100 g,
200 g
10 g, 20 g,
50 g, 100 g,
200 g, 300 g
Dimensions
80 (W) × 230 (D) × 90 (H) mm
Weighing pan
φ34 mm 50 × 50 mm
Weighing
unit
Mass
Approx. 2.3 kg
Dimensions
237 (W) × 150 (D) × 155 (H) mm
Display
Mass
Approx. 2 kg
Connection cable
Approx. 2 m (between weighing unit and display)
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 15VA (supplied to the AC adapter )
Minimum unit mass
0.1 mg
Counting
mode
Number of samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 mg 10.0 mg
Percent
mode
Minimum 100%
display
0.01%、0.1%、1%
(Depends on the reference mass stored.)
OP-01
BCD output, external contact input
(RS-232C and comparator contact output disabled)
OP-07
*
Extension cable 3 m
OP-08
Ethernet interface
(RS-232C and comparator contact output disabled)
OP-19
Stainless steel breeze break(Provided as standard)
Options
OP-20
Metal leveling foot
*1 Specifications when the precision range is used.
*2 With the instrument set to automatic, specifications when measured by loading and removing a 1g mass
repeatedly.
Only one extension cable can be used. The maximum distance between the weighing unit and
the display unit is 5 m (connection cable 2 m + extension cable 3 m).