Instruction manual
18
6. WEIGHING
6-1 Basic Operation (Gram Mode)
1 Plug in the AC adapter.
2 Press the ON:OFF key to display the weighing mode. (The
decimal point position depends on the balance model.)
Notes
The auto display-ON function is available to display the
weighing mode without the key operation when the AC
adapter is plugged in. (Function table: ba5fnc p-on)
The function not to tare at start can be selected. (Function
table: ba5fnc p-tr)
For details on the function table settings, refer to "10.
FUNCTION TABLE".
About the minimum display when starting weighing
With the factory setting, the AD-4212B-23 is 0.01 mg, and the
AD-4212B-101/102 is 0.1 mg. To display 0.001 mg for the
AD-4212B-23, or to display 0.01 mg for the AD-4212B-101/102,
press the SAMPLE key. Then, make sure to warm up the balance
for the appropriate duration.
* In the case of the AD-4212B-23, the displayed unit, when
starting weighing, is mg.
3 Place a container on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The
balance displays 0.0000 g. (The decimal point position
depends on the balance model.)
4 Place a sample on the pan or in the container. Wait for the
stabilization indicator to be displayed and read the value.
5 Remove the sample and container from the pan.
Notes
To use another unit, refer to “5-2 Changing the Unit”.
Press the SAMPLE key to switch the minimum weighing value.
The function to turn on or off the minimum weighing value
automatically when weighing is started can be selected.
(Function table: ba5fnc rng ) For details, refer to "10.
FUNCTION TABLE".
The weighing data can be stored in memory. For details,
refer to “14. DATA MEMORY”.
Weighing
pan
Sample
Remove the
sample
(Software
version
display)
(Display OFF)
(All segments
display)
(Weighing
mode)*
Container