Manual
HV-WP/HW-WP Series Page 27
10. Accumulation Function
Function table Meaning and purpose
A
f8 4
When the display is a stable value and without nearly-zero, the
scale accumulates the data automatically. The next accumulation
can be performed after the display is nearly-zero.
Use Recording the number and mass of articles removed from
the weighing pan. (Put the articles on the weighing pan.
Press the TARE switch at each removal.)
A
A
A
A
A
A
f8 5
At each finish of the full/dribble batch function, the scale accumulates
the data automatically.
Use Packaging articles like a powder, it is used for recording the
bag number and total mass.
A
A
A
10.1. Preparation (Setting Parameters)
Step 1 Turn off the display.
Press the ON/OFF switch while the ZERO switch is
pressed and held. The function table is displayed.
Step 2 Press the ENTER switch to display an item of the
accumulation function ( f8 ).
Step 3 Select a parameter of the accumulation function
( f8 1 ~ f8 4 ) with the ∧ switch.
Step 4 Press the ENTER switch and the F switch to
display end . Press the ENTER switch to return
to the normal weighing mode.