Manual

42
KF format 5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
z This format consists of fourteen characters excluding the terminator.
z This format has no header characters.
z The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
z This format outputs the unit only for a stable value.
MT format 5if type 3
z A header of two characters indicates the balance condition.
z The polarity sign is used only for negative data.
z The weighing data uses spaces in place of the leading zeros.
z The character length of this format changes dependent upon the unit
NU (Numerical) format 5if type 4
This format outputs only numerical data.
z This format consists of nine characters excluding the terminator.
z The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
CSV format 5if type 5
z Separates the data of A&D standard format and the unit by a comma (,).
z Outputs the unit even when the data is overloaded.
z When the ID number, data number, time and date are added, outputs the ID number, data
number, date, time and weighing data in this order and separates each item by a comma and
treats all the items as one group of data.
Note
To add the ID number, data number, time and date, the function settings must be changed.
+ 12.78 g
C
R
L
F
Data
Unit
Terminator
S 12.78 g
C
R
L
F
Header
Unit
TerminatorData