Instruction Manual

Page 16 AD-4406 Weighing Indicator
Press and hold the ON/OFF key and press the ESC key.
No parameters are changed,
CAnCEL is displayed and the calibration mode is finished.
Step18 Press the ON/OFF key to turn the display off.
5.3.
5.3.5.3.
5.3. Weighing Range Function
Weighing Range FunctionWeighing Range Function
Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Note When single range is used, performing this function is not required.
Example 1 The gross display.
Specified parameters:
First range Range = 50.00 kg, division 0.02 kg
Second range Range = 100.00 kg (maximum capacity), division 0.1 kg
Display
0 kg to 50 kg : The first range, division 0.02 kg.
50 kg to 100 kg : The second range, division 0.1 kg.
division 0.02 kg
ޓޓޓdivision 0.1 kg
0 kg 50 kg 100 kg Gross
Weighing value
Example 2 The net display using a 40kg tare value.
Specified parameters: The same parameters as example 1.
Display
-40 kg to 50 kg : The first range, division 0.02 kg.
50 kg to 60 kg : The second range, division 0.1 kg.
division 0.02 kg
division 0.1 kg
-40 kg 0 kg 50 kg 60 kg Net
Tare value 90 kg 100 kg
40 kg Weighing value
5.3. Weighing Range Function